Atomic Absorption Spectroscopy
Measures metal concentrations by absorption of light by free atoms.
Practise Analytical, Lab & Spectroscopy All Analytical, Lab & Spectroscopy topics
Measures metal concentrations by absorption of light by free atoms.
Practise Analytical, Lab & Spectroscopy All Analytical, Lab & Spectroscopy topics
Atomic absorption spectroscopy quantifies metals by converting the sample into free gaseous atoms, usually in a flame or a graphite furnace, and measuring how much light at a characteristic wavelength these atoms absorb. A hollow cathode lamp made from the analyte element emits sharp lines of that element, so the absorption is specific and largely free from interference by other elements. According to the Beer-Lambert relation, absorbance is proportional to the concentration of ground-state atoms, which is in turn proportional to the analyte concentration in the sample. Flame AAS handles concentrations in the ppm range, while graphite furnace AAS reaches ppb levels. Calibration with standards and background correction are essential.
Key ideas: Use matched lamp and wavelength, aspirate blank and standards, build a calibration curve from 0 to about 0.5 absorbance, and read the unknown. Air-acetylene flame (about 2300 C) suits most metals, while refractory elements such as Al and V need nitrous oxide-acetylene (about 2900 C). Chemical interferences, such as phosphate binding calcium, are reduced with releasing agents like lanthanum. Deuterium lamp background correction removes scattering.
Common mistakes and connections: Students confuse AAS with emission methods. In AAS, atoms absorb light from an external lamp, so the sample is not the source. Different elements normally need different lamps, which limits multi-element work. Samples must be dilute and free of solids, and the nebulizer can clog. Related to Beer-Lambert law, flame photometry and ICP techniques.
Atomic absorbance rises with analyte concentration and is calibrated with standards; it curves over at high concentration, so samples are kept in the linear range. The interactive lesson shows this as a live animation or chart.
Lead in drinking water is measured by graphite furnace AAS at 283.3 nm.
Measures metal concentrations by absorption of light by free atoms.
Atomic Absorption Spectroscopy is a advanced topic in Analytical, Lab & Spectroscopy. It builds on beginner and intermediate topics, so review those first if needed.
Study these earlier topics in Analytical, Lab & Spectroscopy first: Integration in NMR, 13C NMR, MRI.
The next topics in Analytical, Lab & Spectroscopy are Flame Photometry and ICP-MS.
Yes. The interactive lesson includes a self-check question on Atomic Absorption Spectroscopy, and the Analytical, Lab & Spectroscopy quiz draws on every topic in this area.
Atomic Absorption Spectroscopy is topic 51 of 99 in Analytical, Lab & Spectroscopy.