Calibration Curve
A graph of signal versus known concentration to determine unknowns.
Practise Analytical, Lab & Spectroscopy All Analytical, Lab & Spectroscopy topics
A graph of signal versus known concentration to determine unknowns.
Practise Analytical, Lab & Spectroscopy All Analytical, Lab & Spectroscopy topics
A calibration curve is a graph of instrument response, such as absorbance or peak area, plotted against the known concentrations of a series of standards, and it is used to convert the response of an unknown into a concentration. A linear least-squares fit gives a slope and intercept, and the unknown concentration is found by reading its response off the line. The standards should span the expected concentration range and include a blank, and the unknown should fall within the calibrated range to avoid extrapolation. A coefficient of determination R squared close to 1 indicates good linearity, but residuals should also be examined. The calibration should be checked regularly because instrument drift changes the slope.
Key ideas: Fit the line y = mx + b, then x = (y minus b)/m for the unknown. Use at least five standards across the range, run them in random order, and replicate. The standard error of the estimate gives the uncertainty. For complex matrices use the standard addition method to cancel matrix effects. Detection limit comes from the slope and blank noise: 3 s/m.
Common mistakes and connections: Students force the line through the origin without testing the intercept, extrapolate beyond the top standard where curvature occurs, or plot concentration on the y-axis. High R squared does not guarantee accuracy, since a systematic bias can remain. Standards prepared by serial dilution propagate any early error. Links to Beer-Lambert limits, error analysis and detection limit.
A straight line through standards converts a measured absorbance of 0.42 into a concentration; the unknown must lie inside the calibrated range. The interactive lesson shows this as a live animation or chart.
An unknown with absorbance 0.42 is read off a line drawn through standards from 0.1 to 1.0 mg/L.
A graph of signal versus known concentration to determine unknowns.
Calibration Curve is a intermediate topic in Analytical, Lab & Spectroscopy. It builds on the vocabulary of beginner topics in this area.
Study these earlier topics in Analytical, Lab & Spectroscopy first: UV-Visible Spectroscopy, Beer-Lambert Law, Colorimetry.
The next topics in Analytical, Lab & Spectroscopy are Infrared Spectroscopy and Characteristic IR Absorptions.
Yes. The interactive lesson includes a self-check question on Calibration Curve, and the Analytical, Lab & Spectroscopy quiz draws on every topic in this area.
Calibration Curve is topic 40 of 99 in Analytical, Lab & Spectroscopy.