Detection Limit
The lowest concentration reliably distinguished from background.
Practise Analytical, Lab & Spectroscopy All Analytical, Lab & Spectroscopy topics
The lowest concentration reliably distinguished from background.
Practise Analytical, Lab & Spectroscopy All Analytical, Lab & Spectroscopy topics
The detection limit is the lowest analyte concentration that a method can reliably distinguish from a blank, and it defines the sensitivity of the method. A common definition sets the limit of detection at three times the standard deviation of the blank divided by the slope of the calibration curve, while the limit of quantification is taken as ten times that quantity. Results below the detection limit are reported as not detected, because their signal cannot be separated from noise. The limit depends on instrument noise, the calibration slope and the sample matrix, and should be determined experimentally. It sets the practical lower bound for environmental and clinical analysis.
Key ideas: LOD = 3 s_blank/m and LOQ = 10 s_blank/m, where m is calibration slope. Estimate s_blank from at least seven replicate blanks or low-level spikes. Between LOD and LOQ a result is detected but cannot be quoted precisely. Concentration units such as ppm (mg/L), ppb (microgram/L) and ppt (ng/L) must be tracked. A method with a steeper slope has a lower detection limit for the same noise.
Common mistakes and connections: Students confuse detection limit with sensitivity, which is the slope, or with the lowest standard they measured. Reporting zero for results below LOD is misleading, so state less than the LOD. Contamination raises the blank and the LOD. Instrument detection limit is lower than method detection limit, which includes sample preparation. Related to calibration, standard deviation and ICP-MS.
With a blank standard deviation of 0.5 counts, a signal three standard deviations above the blank (1.5 counts) is the smallest reliably distinguishable from noise. The interactive lesson shows this as a live animation or chart.
An ICP-MS method with a blank standard deviation of 0.5 counts and a slope of 100 counts per ppt has a detection limit of 0.015 ppt.
The lowest concentration reliably distinguished from background.
Detection Limit is a advanced topic in Analytical, Lab & Spectroscopy. It builds on beginner and intermediate topics, so review those first if needed.
Study these earlier topics in Analytical, Lab & Spectroscopy first: Conductometric Titration, Error Analysis, Standard Deviation.
The next topics in Analytical, Lab & Spectroscopy are Sampling and Quality Control.
Yes. The interactive lesson includes a self-check question on Detection Limit, and the Analytical, Lab & Spectroscopy quiz draws on every topic in this area.
Detection Limit is topic 66 of 99 in Analytical, Lab & Spectroscopy.